TVM 5215AC Tochline

Micro Vickers Hardness Tester

The micro-Vickers hardness tester TVM 5215AC is designed to measure the hardness of metals and alloys according to the Vickers method according to GOST R 8.695-2009 (ISO 6507-1: 2005),
GOST 2999-75

Micro-Vickers hardness tester TV 5215AC is a stationary measuring instrument consisting of a load application system and a measuring unit.
The principle of operation of the hardness tester is based on the static indentation of the diamond pyramidal tip with subsequent measurement of the lengths of the diagonals of the reconstructed print.
The device has an automatic rotary turret, an electromechanical loading drive, an electronic load monitoring system, and an integrated optical system.
When measured according to the Vickers method, the system for applying the load of devices provides the application of loads of 0.098; 0, 245; 0.49; 0.98; 1.96; 2.94; 4.9; 9.8; 19.6N
The design of the hardness tester is reliably protected from unauthorized access, adjustment of factory settings and software.
Special software for the built-in microprocessor of the hardness tester software is used to enter the initial parameters and start the load application cycle, record, store and statistically process the measurement results and display them on the digital liquid crystal display.

Features of the digital-to-digital unit:
• Interactive work;
• Hardness scale selection;
• Setting the exposure time under load;
• Display of loading process and holding time;
• Sorting into hardness groups LESS, NORMAL, MORE;
• Statistical processing of a series of measurements;
• Saving results in the memory of the Central Bank;
• Auto-tuning (calibration) according to standard measures of hardness;
• Recalculation of hardness values ​​in other scales and methods;
• Touch control;
• Multi-user interface;
• Semi-automatic optical system for measuring the size of the print according to the Vickers method with automatic data transfer and obtaining hardness on the display.

SPECIFICATIONS
Test loads on Vickers scales, N: 0,098; 0, 245; 0.49; 0.98; 1.96; 2.94; 4.9; 9.8; 19.6
Vickers hardness measurement range, HV: from 8 to 3000
Limits of permissible relative error in measuring hardness according to Vickers scales, %: ±3
Limits of permissible absolute measurement error of a measuring microscope, mm: ±0,001
Dimensions X-Y coordinate table, mm: 100х100
X-Y stroke of the coordinate table, mm: 25х25
The maximum height of the sample, mm: 170
Depth of a working zone, mm: 115
Overall dimensions (length x width x height), mm, no more than: 490x185x515
Weight, kg, no more than: 43